Sign in →

Test Code TIWB Titanium, Blood


Specimen Required


Patient Preparation: High concentrations of gadolinium and iodine are known to interfere with most metal tests. If either gadolinium- or iodine-containing contrast media has been administered, the specimen should not be collected for at least 96 hours.

Supplies: Metal Free B-D Tube (EDTA), 6 mL (T183)

Collection Container/Tube: Plain, royal blue-top (EDTA) plastic trace element blood collection tube

Specimen Volume: 1.0 mL

Collection Instructions:

1. See Metals Analysis Specimen Collection and Transport for complete instructions.

2. Send whole blood specimen in original collection tube. Do not aliquot.


Secondary ID

614612

Useful For

Detecting and monitoring titanium exposure and potential implant status in patients with orthopedic implants

Method Name

Triple-Quadrupole Inductively Coupled Plasma Mass Spectrometry (ICP-MS/MS)

Reporting Name

Titanium, B

Specimen Type

Whole blood

Specimen Minimum Volume

0.3 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Whole blood Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Reject Due To

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK

Clinical Information

Titanium is the ninth most abundant element in the earth's crust. Its light weight and high strength are useful in alloys for diverse applications. There is no evidence that titanium is an essential element. Due in part to titanium's oxide formation propensity, the element is considered to be nontoxic. Soils, drinking water, and air all contain trace amounts of titanium. The food processing industry uses large quantities of titanium as a food additive; processed foods contain higher levels than are found in most produce and organic food products. The average daily oral intake through food consumption is 0.1 to 1 mg/day, which accounts for more than 99% of exposure. Gastrointestinal absorption of titanium is low (approximately 3%), and the majority of ingested titanium is rapidly excreted in the urine and stool. The total body burden of titanium is usually in the range of 9 to 15 mg, a significant portion of which is contained in the lungs. Titanium dust entering the respiratory tract is nonirritating and is almost completely nonfibrogenic in humans.

 

Titanium-containing alloys are used in some artificial joints, prosthetic devices, and implants. Titanium dioxide allows osseointegration between an artificial medical implant and bone. Despite their wide use, exposure to these materials has not been linked to toxicity. In one study, patients monitored up to 36 months following joint replacement with titanium-containing joints showed a statistically significant increase in detectable titanium. While titanium concentrations are not a measure of toxicity, they are useful in determining whether implant breakdown is occurring. Blood titanium concentrations are likely to be increased above the reference range in patients with metallic joint prosthesis. Prosthetic devices produced by Zimmer Company and Johnson and Johnson typically are made of aluminum, vanadium, and titanium. This list of products is incomplete, and these products change occasionally; see prosthesis product information for each device for composition details.

Reference Values

0-17 years: Not established

≥18 years: <2 ng/mL

Interpretation

Clinically, this test is used to detect and monitor titanium exposure and potential implant status in patients with orthopedic implants. Increased concentrations in blood have been proposed as a marker of component wear.

CPT Code Information

83018

LOINC Code Information

Test ID Test Order Name Order LOINC Value
TIWB Titanium, B 59753-4

 

Result ID Test Result Name Result LOINC Value
614612 Titanium, B 59753-4