Test Code QUIN1 Quinidine, Serum or Plasma

Important Note

Do not use a gel-barrier tube. The use of gel-barrier tubes is not recommended due to slow absorption of the drug by the gel. Depending on the specimen volume and storage time, the decrease in drug level due to absorption may be clinically significant.

Useful For

This Class 1A drug is useful in both supraventricular and ventricular arrhythmias. It major uses are to maintain sinus rhythm after conversion of atrial flutter or fibrillation, to prevent ventricular tachycardia, and for long-term prophylaxis in patients with AV nodal reentrant tachycardia and automatic atrial tachycardia. Quinidine also has been used to prevent symptomatic premature supraventricular and ventricular complexes. Because it slows conduction and prolongs the refractory period of the accessory pathway and suppresses automaticity of ectopic pacemakers, quinidine may prevent recurrences of paroxysmal supraventricular tachycardia caused by reentry over a concealed pathway or AV reciprocating tachycardia associated with the Wolff-Parkinson-White syndrome. It also may slow the ventricular response to atrial flutter or fibrillation in the preëxcitation syndrome. Quinidine is often preferred to procainamide for long-term therapy because elevated antinuclear antibody titers and drug-induced lupus are common during prolonged therapy with procainamide.

Specimen Information

Obtain specimen using standard phlebotomy techniques.

No special patient preparation is necessary.

 

Container Specimen Temperature Collect Vol Submit Vol Min Vol Stability
Red Top Serum Room Temp 4 mL 1 mL 0.5 mL 14 Days
Grn Top (No Gel) Plasma Room Temp 4 mL 1 mL 0.5 mL 14 Days

 

 

Reject Due To:

- Moderate or Marked Hemolysis

- Specimen unlabeled or mislabeled

- Gel barrier tube used for collection

 

 

Instrumentation / Method

Immunoassay

Test Schedule / Analytical Time / Test Priority

Daily / 1 - 3 days / Not Available STAT

Reference Range

Male and Female: 

Therapeutic Range: 2.0 - 5.0 ug/mL

 

CPT / LOINC / Order Code

Meditech Medent LOINC CPT
QUIN1 QUIN1 6694-4 80194