Test Code HIVC HIV-1 & 2 Antigen and Antibody, 4th Generation
Specimen Information
Obtain specimen using standard phlebotomy techniques.
If taking a suppliment containing Biotin, patients should be instructed to discontinue its use 12 hours before blood collection.
Container | Specimen | Temperature | Collect Vol | Submit Vol | Min Vol | Stability |
SST | Serum | Refrigerate | 4.0 mL | 2.0 mL | 1.0 mL | 7 days |
*Lithium Heparin | Plasma | Refrigerate | 4.0 mL | 2.0 mL | 1.0 mL | 7 days |
Reject Due To:
- Moderate or Marked Hemolysis
- Specimen unlabeled or mislabeled
* Lithium Heparin is an acceptable sample type for testing HIV, however it is NOT acceptable for reflex HIV Antibody Confirmation / Differentiation testing.
Instrumentation / Method
Ortho Vitros XT 7600 / Immunometric
Test Schedule / Analytical Time / Test Priority
Daily / Same Day / Available STAT
Reference Range
Qualitative results are based on a cutoff value of 1.00.
<1.00 (Negative) | Sample is negative for anti-HIV-1 (including Group O), anti-HIV-2, and p24 antigen. |
≥1.00 (Reactive) | Sample is reactive for anti-HIV-1 (including Group O), anti-HIV-2, and/or p24 antigen. |
CPT / LOINC / Order Code
Meditech | Medent | LOINC | CPT |
HIVC | HIVC | 31201-7 | 87389 |