Sign in →

Test Code CL Chloride

Useful For

Assessing water balance and osmotic pressure in conjunction with Sodium.

Specimen Information

Obtain specimen using standard phlebotomy techniques.

No special patient preparation is necessary.

 

Container Specimen Temperature Collect Vol Submit Vol Min Vol Stability
SST Serum Refrigerate 4.0 mL 1.0 mL 0.6 mL 7 days
Lithium Heparin Plasma Refrigerate 4.0 mL 1.0 mL 0.6 mL 7 days

Reject Due To:

- Moderate or Marked Hemolysis

- Specimen unlabeled or mislabeled

Reference Range

Male and Female: 98 - 107 mmol/L

 

Critical Value(s): <80 or >120 mmol/L

Test Schedule / Analytical Time / Test Priority

Daily / Same Day / Available STAT

Instrumentation / Method

Ortho Vitros XT 7600 / Ion Selective Electrode

CPT / LOINC / Order Code

Meditech Medent LOINC CPT
CLS CLS 2075-0 82435