Test Code CBART Blood Gas, Cord, Arterial
Specimen Information
Container | Specimen | Storage | Stability |
Heparinized Syringe | Whole Blood, Cord | Ambient | 30 minutes |
Refer to the Umbilical Cord Bloos Gas Sampling policy for collection instructions.
After collection, firmly attach the cap (Filter-Pro) onto the syringe, While holding the luer end up, tap the syringe to move any air bubbles to the top. To remove the air, SLOWLY press the plunger and continue to advance until the sample wets the filter.
Reject Due To:
- Specimens received with needles attached
- Evidence of clotting or hemolysis
- Specimens collected in vacutainer tubes
- Specimens with excessive air bubbles
- Specimens received greater than 30 minutes after collection
- Non-heparinized samples
- Other sample types such as urine, CSF, pleural fluid
- Evidence of anticoagulant or IV contamination
- Unlabeled/Mislabeled specimens
- Specimens received on ice / cold packs
Instrumentation / Method
pH - iSTAT / Ion Selective Electrode Potentiometry at 37 C
pO2 - iSTAT / Amperometrically
pCO2 - iSTAT / Direct Potentiometry at 37 C
HCO3 - iSTAT / Calulated from pH and pCO2 results
Test Schedule / Analytical Time / Test Priority
Daily / Immediate / Available STAT
Reference Range
No reference ranges have been established for this test.
LOINC / CPT/ Ordering
Analyte | CPT | LOINC |
CBART | 82803 | |
pH, Arterial, Cord | 28646-8 | |
pO2, Arterial, Cord | 28648-4 | |
pCO2, Arterial, Cord | 28644-3 | |
HCO3, Arterial, Cord | 28640-1 | |
BE / BD, Arterial, Cord | N/A |